Dictionary / Measurement Geometry

Axial Measurement

Measurement Geometry
Field strength measured along the central axis of the applicator coil, perpendicular to its face. Axial readings represent the maximum field value at a given distance and are used as the reference standard for PEMF device characterization. Field strength falls off rapidly with distance. Values in the report are at 0mm (contact surface) unless otherwise noted.
A top-down scan grid over a round accessoryA round accessory seen from above with measurement points stepped outward from the center along four axes at a fixed spacing. A couple of points sit past the accessory edge to capture the field as it falls off.Point spacingOrigin (center)Accessory edgeX axisY axisExtra points beyond the edgeThe engineer measures the field twice as far apart once past the accessory edge.© 2026 Gauss Labs
The field is measured on a grid. Starting at the center, the test engineer steps a probe outward along each axis at a fixed point spacing, taking a reading at every point, and continues a little past the accessory edge to catch the field as it fades. The origin is the center of the accessory surface.
measurement